To all potential bidders for Procurement ICB # 1.2.2.1
- Date of the news:
- Component:
- Subproject:
Serving People Improving Health Project
Loan No.: 8475- UA dated March 19, 2015
Subproject: “Support to Health Care System Reform of Dnipropetrovsk oblast”
Procurement: No. ICB – 1.2.2.1 Medical Equipment: angiograph to equip Municipal Enterprise "Dnipropetrovsk oblast clinical center for cardiology and cardiosurgery of DRC”
ANNEX NO. 4
The provisions of this Addendum are being issued to modify or amend certain provisions of the Bidding Documents as a result of changing circumstances or additional information being available from the Employer or in response to the queries raised by the prospective bidders at the pre-bid meeting or queries raised by the prospective bidders in writing. The Bidding Documents are hereby modified as below.
Modification No. |
Bidding Document Page No. |
Bidding Document Reference Clause |
Clause in the Original Bidding Document |
Modified Clause |
---|---|---|---|---|
1 |
64 |
8.1.1.4. |
C-arm depth not less than 105 cm. |
C-arm depth not less than 89 cm. |
2 |
65 |
8.1.3.7. |
Quantum efficiency of detector not less than 80 % |
Quantum efficiency of detector not less than 75 % |
3 |
65 |
8.1.4.5. |
Maximum heat capacity of X-ray tube anode not less than 3 000 000 thermal units |
Maximum heat capacity of X-ray tube anode not less than 2,400,000 thermal units |
4 |
65 |
8.1.4.6. |
Cooling rate of anode not less than 560,000 thermal units per minute |
Cooling rate of anode not less than 500,000 thermal units per minute |
5 |
65 |
8.1.4.7. |
Maximum heat capacity of X-ray tube unit not less than 6,500,000 thermal units |
Maximum heat capacity of X-ray tube unit not less than 5,100,000 thermal units |
6 |
65 |
8.1.4.9. |
Number of focal spots of the tube not less than 3 |
Number of focal spots of the tube not less than 2 |
7 |
65 |
8.1.4.10. |
Minimum size of focal spot not more than of 0,3 mm |
Minimum size of focal spot not more than of 0.6 mm |
8 |
66 |
8.1.5.11. |
Integrated CD / DVD recorder for archiving DICOM image files together with DICOM viewer program |
Integrated or independent CD/DVD recorder for archiving DICOM image files together with DICOM viewer program |
9 |
67 |
8.1.7.11. |
Availability of the permit from the MoH of Ukraine for use in medical practice (a copy of the Certificate on State registration of medical designation product with appendices or a list of accessories to it at the time of delivery to be attached) |
“taking into account requirements of Decrees of the Cabinet of Ministers of Ukraine dated October 02, 2013 No. No. 753*, №754**, №755***, it is necessary to furnish the Certificate of Conformity to “Technical regulation on medical designation devices” or guarantee letter from the Bidder that copies of documents to the offered Goods would be furnished prior to remitting the advance payment. * - Decree of the Cabinet of Ministers of Ukraine “On Approval of the Technical regulation on medical designation devices” dated October 02, 2013 No. 753. ** - Resolution of the Cabinet of Ministers of Ukraine "On Approval of Technical regulation on medical designation devices for the diagnosis in vitro” dated October 02, 2013 No. 754. *** - Resolution of the Cabinet of Ministers of Ukraine "On Approval of Technical Regulation on the active medical devices that are implanted” dated October 02, 2013 No. 755. |
Please, confirm the reception of this message to the following e-mail address: dnipro_project@ukr.net.
Director of Health Care Department
of Dnipropetrovsk oblast state administration N.Y. BUDYAK